Skip to content
Skip to content
  • News
  • Events
  • eBooks
  • Blog
  • Careers
  • Contact
  • News
  • Events
  • eBooks
  • Blog
  • Careers
  • Contact
KNOWLEDGEBASE
  • About Us
  • Products

    Fluorescence Spectrometers

    • FLS1000 Photoluminescence Spectrometer
    • FS5 Spectrofluorometer
    • LifeSpec II Lifetime Spectrometer
    • Mini-tau Lifetime Spectrometer

    Raman Microscopes

    • RM5 Raman Microscope
    • RMS1000 Multimodal Confocal Microscope

    Transient Absorption

    • LP980 Transient Absorption Spectrometer

    FTIR Spectrometers

    • IR5 FTIR Spectrometer

    Lasers and LEDs

    • Pulsed Lasers
    • Gas Lasers
    • Customisation Options
    View All Products
  • Techniques
  • Applications
KNOWLEDGEBASE
Edit Content
  • About Us
  • Products
  • Techniques
  • Applications
  • Knowledgebase
  • eBooks
  • News
  • Events
  • Blog
  • Careers
  • Contact Us

RESOURCES

Raman Thermometry of a Silicon Semiconductor Chip

  • March 13, 2025

KEY POINTS

  • The LO phonon mode in silicon exhibits a characteristic change in Raman shift with temperature, decreasing in energy as temperature increases because of thermal expansion of the crystal lattice.
  • Calibration of the LO phonon mode Raman shift allows for non-invasive and localised temperature measurements in silicon samples.
  • The strong linear correlation between peak position and temperature enables accurate thermometry of silicon devices.
Edit Content

Introduction 

Temperature-dependent (TD) Raman microscopy is a powerful and non-destructive method that can probe the thermal and electronic properties of semiconductor materials with sub-micron spatial resolution.  

Higher temperatures influence crucial properties of semiconductors, leading to increased conductivity and reduced resistance.1 These properties are intimately linked to the performance of the final device. It is, therefore, essential to have sensitive and specific techniques that can be used to non-invasively investigate the temperature of devices to rationalise the observed performance.  

TD Raman microscopy is capable of this because the Raman spectrum of a semiconductor material is highly sensitive to changes in temperature.2 More specifically, shifts in the peak position of the phonon modes in a semiconductor can be directly related to thermal expansion or contraction within the crystal lattice, and calibrations can be performed that directly link this to temperature. This technique is also known as Raman thermometry. 

In this Application Note, we demonstrate the power of TD Raman microscopy by investigating the temperature-induced phonon mode shifts in a silicon (Si) chip with an Edinburgh Instruments RM5 Confocal Raman Microscope equipped with a temperature stage. We show that Raman thermometry can be used to non-invasively determine the temperature of Si-based devices.  

 

Materials and Methods

The sample analysed was a 10 mm x 10 mm chip from an n-doped Si wafer. The chip was placed in a Linkam THMS600 Heating and Cooling Stage, which allows for temperature control in a range between 78 K and 873 K.  

The stage was fixed underneath the microscope of an RM5 equipped with a 50X 0.5 NA long-working distance objective, a 532 nm laser and a 2400 gr/mm diffraction grating, Figure 1, which allowed for very high spectral resolution to be achieved. Spectral band positions were calculated using a twice-smoothed Lorentzian function in Ramacle® software. The temperature of the stage was controlled using Ramacle, and spectra were recorded between 173 K and 473 K in 25 K steps.  

 

Figure 1. An n-doped Si wafer inside a Linkam THM600 fitted to an RM5. 

Raman Thermometry: Calibrating the Si LO Phonon Peak Position to Temperature 

Raman thermometry is the determination of the temperature of a sample using features of its Raman spectrum. The technique relies on the spectrometer having sufficiently high spectral resolution. This could potentially be a useful technique in a variety of applications, including microelectronics, materials science, and biomedical sensing. To achieve this, a calibration curve of Raman shift versus temperature is required. This is produced from spectra recorded at the same point of the sample at various known temperatures.  

Figure 2 shows Raman spectra recorded on an n-doped Si chip between 173 K and 473 K in 25 K steps. These spectra show that as the temperature increases, the intensity of the dominant LO phonon band in Si (ca. 520.5 cm-1 at room temperature) decreases, and the position of the band gradually shifts to a lower wavenumber.  

 

Figure 2. Raman spectra recorded from an n-doped Si chip between 173 K and 473 K in 25K steps.  

The decrease in intensity of the LO phonon mode with increasing temperature can be understood using the Boltzmann distribution, Equation 1, which describes the statistical population of the ground and excited vibration energy levels relative to temperature. 

Nn and Nm are the number of molecules in the excited and ground vibrational energy levels, g describes the degeneracy of the energy levels and for the purposes of this work can be ignored, (En–Em) is the energy difference (kg m2 s-2) between the excited and ground vibrational energy levels, k is the Boltzmann constant (1.38065×10-23 m2 kg s-2), and T is the temperature of the sample. To put this equation simply, the ratio of excited state to ground state populations is proportional to temperature.3  

The spectra in Figure 2 show the Stokes Raman LO phonon band, which always has a positive Raman shift relative to the Rayleigh scattering line of the laser at 0 cm-1. Stokes Raman scattering can only occur when the material under investigation is in the ground vibrational level prior to optical excitation, and the Boltzmann distribution describes the decreasing population of the ground state as the sample temperature increases. It therefore follows that as the sample temperature increases, the Stokes intensity decreases because the sample is raised to a higher vibrational energy state by thermal excitation, Figure 3. Conversely, using the same logic, the intensity of anti-Stokes scattering will increase at higher temperatures.  

 

Figure 3. A Jablonski diagram showing the effect of (A) low and (B) high temperature on the population of ground and vibrational states, and the resulting Stokes and anti-Stokes Raman scattering intensity.  

Tracking the ratio of anti-Stokes to Stokes scattering with increasing temperature for the same band is a potential method for determining the temperature of a sample. It requires the use of a notch filter in the microscope, in order to detect both Stokes and Anti-Stokes signals.4 In this work, the RM5 microscope used was equipped with an edge filter, which does not allow for anti-Stokes Raman scattering to be observed but enables lower-wavenumber detection than a notch filter. However, this is not the only method for determining the temperature of a sample with Raman microscopy.  

The position of the LO phonon band changes with increasing temperature because of thermal expansion, which causes Si-Si bond lengths within the crystalline lattice to increase. The lengthening of the bond away from its equilibrium position decreases the energy of the bond. This leads to a decrease in the Raman shift of the LO phonon mode because it is proportional to the bonding energy within the lattice.5  This is why we observe the Raman shift of the LO phonon mode gradually decreasing between 173 K and 473 K. This phenomenon is only useful for temperature prediction if a quantitative relationship exists.  

Figure 4 shows the peak position of the LO phonon mode plotted against the temperature. A strong linear relationship is observed between the two variables (R² = 0.9972). In a controlled environment with a defect-free, strain-free Si sample, the LO phonon band position can serve as a potential temperature indicator using simple linear regression. This level of accuracy is possible because of the extremely high spectral resolution achieved when using a 532 nm laser with a 2400 gr/mm diffraction grating.  

Figure 4. A plot showing the LO phonon peak position for an n-doped Si chip against temperature. The linear regression reveals a strong correlation between the two variables.  

Conclusion

TD Raman microscopy is a powerful tool for probing the thermal and electronic properties of semiconductors. By monitoring the LO phonon mode in an n-doped Si chip from 173 K to 473 K, we demonstrated the ability to determine its temperature non-invasively.  

 

References

  1. V. K. Khanna, in Extreme Temperature and Harsh-Environment Electronics: Physics, Technology and Applications, IOP Publishing, 2017. 
  1. T. R. Hart et al., Phys. Rev. B, 1970, 1, 638-642.  
  1. L. S. Ornstein et al., Phys. Rev., 1929, 34, 720.  
  1. B. Gardner et al., Anal. Chem., 2016, 88, 832-837. 
  1. L. Deng et al., Appl. Phys. Lett., 2014, 104, 051907.  

RELATED PRODUCTS

RMS1000

Multimodal Confocal Microscope

VIEW

RM5

Raman Microscope

VIEW

Contact our expert team today to find out more about how our products can improve your research

Contact Us
PrevPrevious
NextNext
Previous Fluorescence Lifetime Imaging Microscopy (FLIM) Analysis of Convallaria Next Circularly Polarised Luminescence from TbPO4·nH2O Nanocrystals

RESOURCES

Tags:
  • Application Notes
  • Semiconductors
  • RM5
  • RMS1000
  • Raman
Application Note
Download PDF
Featured Product:

RM5 Raman Microscope

View more »
Suggested Reading:

Raman Imaging of Strain in a Silicon Semiconductor Wafer

View more »
Share:

Keep up to date with the latest from Edinburgh Instruments

Join our mailing list and keep up with our latest videos, app notes and more!

LOCATION:
  • Edinburgh Instruments Ltd.
    2 Bain Square, Kirkton Campus, Livingston, EH54 7DQ.
  • sales@edinst.com
  • +44 1506 425 300
ABOUT:
  • About Us
  • Techniques
  • Applications
  • Knowledgebase
  • About Us
  • Techniques
  • Applications
  • Knowledgebase
PRODUCTS:
  • Fluorescence Spectrometers
  • Raman Microscopes
  • UV-Vis Spectrophotometers
  • Transient Absorption
  • FTIR Spectrometers
  • Lasers and LEDs
  • Customisation Options
  • Software
  • Upgrades
  • All Products
  • Fluorescence Spectrometers
  • Raman Microscopes
  • UV-Vis Spectrophotometers
  • Transient Absorption
  • FTIR Spectrometers
  • Lasers and LEDs
  • Customisation Options
  • Software
  • Upgrades
  • All Products
LEGALS:
  • News
  • Events
  • Blog
  • Careers
  • Contact Us
  • Terms and Conditions
  • Privacy Policy
  • News
  • Events
  • Blog
  • Careers
  • Contact Us
  • Terms and Conditions
  • Privacy Policy
SOCIALS:
Youtube Linkedin X-twitter Facebook
©2024 Edinburgh Instruments. Registered in England and Wales No: 962331. VAT No: GB 271 7379 37
Manage Consent
To provide the best experiences, we use technologies like cookies to store and/or access device information. Consenting to these technologies will allow us to process data such as browsing behaviour or unique IDs on this site. Not consenting or withdrawing consent, may adversely affect certain features and functions.
Functional Always active
The technical storage or access is strictly necessary for the legitimate purpose of enabling the use of a specific service explicitly requested by the subscriber or user, or for the sole purpose of carrying out the transmission of a communication over an electronic communications network.
Preferences
The technical storage or access is necessary for the legitimate purpose of storing preferences that are not requested by the subscriber or user.
Statistics
The technical storage or access that is used exclusively for statistical purposes. The technical storage or access that is used exclusively for anonymous statistical purposes. Without a subpoena, voluntary compliance on the part of your Internet Service Provider, or additional records from a third party, information stored or retrieved for this purpose alone cannot usually be used to identify you.
Marketing
The technical storage or access is required to create user profiles to send advertising, or to track the user on a website or across several websites for similar marketing purposes.
Manage options Manage services Manage {vendor_count} vendors Read more about these purposes
View preferences
{title} {title} {title}