Skip to content
  • News
  • Events
  • eBooks
  • Blog
  • Careers
  • Contact
  • News
  • Events
  • eBooks
  • Blog
  • Careers
  • Contact
KNOWLEDGEBASE
  • About Us
  • Products

    Fluorescence Spectrometers:

    • FLS1000 Photoluminescence Spectrometer
    • FS5 Spectrofluorometer
    • LifeSpec II Lifetime Spectrometer
    • Mini-tau Lifetime Spectrometer

    Raman Microscopes:

    • RM5 Raman Microscope
    • RMS1000 Multimodal Confocal Microscope

    Transient Absorption:

    • LP980 Transient Absorption Spectrometer

    FTIR Spectrometers:

    • IR5 FTIR Spectrometer

    Lasers and LEDs:

    • Pulsed Lasers
    • Gas Lasers

    Options:

    • Upgrades
    • Customisation
    View All Products
  • Techniques
  • Applications
KNOWLEDGEBASE
  • About Edinburgh Instruments
  • Products
  • Techniques
  • Applications
  • Knowledgebase
  • eBooks
  • News
  • Events
  • Blog
  • Careers
  • Contact Us

RESOURCES

Choosing Detectors for your FLS1000 Photoluminescence Spectrometer

  • April 13, 2026

KEY POINTS

  • Choose a detector that covers the entire emission range of your sample while providing sufficient sensitivity for weak signals. The PMT-900 is standard for the visible spectral region, while options such as the PMT-980, PMT-1400, and PMT-1700 extend coverage into the near-infrared region.
  • To perform PLQY, the chosen detector(s) must fully encompass both the excitation and emission spectral regions of your samples.
  • The detector’s transit time spread determines the shortest measurable lifetime. Use a high-speed or high-speed hybrid photodetector for ultrafast decays in the visible range. For the NIR range, note that higher dark-count rates may affect the accuracy of lifetime measurements.

In this Technical Note, we guide you through selecting the most suitable detector configuration for the FLS1000 Photoluminescence Spectrometer. It outlines the key considerations when choosing between photon-counting photomultiplier tubes (PMTs) and analogue detectors, based on sample characteristics and measurement requirements. Readers with a particular interest in near-infrared (NIR) detection are encouraged to consult the companion Technical Note, “Near Infrared PMT Options for the FLS1000 Photoluminescence Spectrometer”. 

Detector Types 

The FLS1000 can accommodate various PMTs and analogue detectors. PMTs are the default option for photoluminescence in the visible spectral range, while NIR detection can be performed with photon-counting or analogue detectors. Some key features to consider when selecting your optimal detector module include: 

Photon-Counting PMTs 

  • Detect individual photons. 
  • Provide a high signal-to-noise ratio (SNR). 
  • Are ideal for weakly emitting samples and time-resolved measurements. 
  • Have a fast time response. 
  • Eliminate noise by photon counting. 

Analogue Detectors (e.g., InGaAs, InAs, InSb): 

  • Measure current directly rather than counting photons. 
  • Can be paired with lock-in amplifiers to enhance weak signals. 
  • Are essential for measurements beyond 1700 nm, although they have lower SNR when compared with PMTs. 
  • Can be used for steady-state and/or time-resolved measurements using Edinburgh Instruments Transient Digitiser (TDS) electronics, which eliminate the need for an oscilloscope. 

 

What is the Spectral Emission Range of Your Sample? 

The most important factor when choosing a detector is whether it can cover the full emission range of your sample. Each PMT has a specific spectral sensitivity window, and measurements are only possible within this range. The ability of a PMT to detect light and convert photons into an electrical signal is measured by its quantum efficiency (QE). This conversion occurs at the PMT’s photocathode, where incoming photons produce photoelectrons. These photoelectrons are then accelerated and amplified by the PMT’s dynodes. The QE at a specific wavelength is usually given as a percentage and is defined as: 

The PMT-900 is the standard option for visible-range detection. It is suitable for most samples emitting in the 200 nm – 870 nm region and offers excellent performance for general photoluminescence applications. If the sample emission extends to ~950 nm, the PMT-900 can be replaced with the PMT-980, which has extended spectral coverage. The QE curves for both detectors are shown in Fig. 1. 

Figure 1. QE curves of the PMT-900 and -980. 

If your sample emission extends further into the NIR or mid-infrared (MIR) region, additional detectors are required. You can choose from the available PMTs, either liquid nitrogen- or thermoelectrically (LN2 or TE)-cooled, or the analogue detector options listed in Tables 1 and 2, respectively. These detectors complement the visible PMTs and can be used as 2nd or 3rd detectors on the same FLS1000 system. 

Table 1. Available PMT options for the visible and NIR regions, along with their corresponding spectral ranges.

Photon Counting DetectorRange
PMT-900200 nm – 870 nm
PMT-980200 nm – 980 nm
PMT-1010200 nm – 1010 nm
PMT-1400-LN2500 nm – 1400 nm
PMT-1400-TE930 nm – 1400 nm
PMT-1700-LN2500 nm – 1700 nm
PMT-1700-TE930 nm – 1700 nm

Table 2. Available analogue detector options for the NIR and MIR regions, along with their corresponding spectral ranges. 

Analogue DetectorRange
InGaAs-1650-TE800 nm – 1650 nm
InGaAs-2050-TE900 nm – 2050 nm
InGaAs-2550-TE900 nm – 2550 nm
InAs-3100-LN21200 nm – 3100 nm
InSb-5500-LN21500 nm – 5500 nm

Important considerations that should be kept in mind:  

  • The sensitivity of any detector declines at the edges of its spectral range. Therefore, although a detector technically reaches a specific wavelength, its performance drops significantly near that limit. 
  • The groove density of the diffraction grating used determines the spectral resolution; typically, a higher density enhances resolution but reduces spectral range. 

 

How Bright Are Your Samples? 

Not all samples emit light with the same intensity. Some produce very weak photoluminescence, particularly in the NIR/MIR regions. The detector’s sensitivity is crucial in these instances. A measure of the detector’s sensitivity that incorporates QE, dark count rate, and noise-equivalent power (NEP) is the number of photons per second required to achieve a signal-to-noise ratio (SNR) of 1. That is: 

where λ is the wavelength of light incident on the detector, h is Planck’s constant and c is the speed of light in vacuum. 

Figures 2 and 3 present a sensitivity comparison using Eq. 2 for all the available FLS1000 detectors. Figure 2 shows a sensitivity comparison between the available PMTs up to 1700 nm, and Figure 3 shows a comparison between the analogue detectors up to 5.5 μm. Note that the higher the value on the y-axis, the less sensitive the detector is, as this figure is the number of photons per second required to achieve an SNR of 1. The figure shows that PMTs can be up to 1,000,000 times more sensitive than analogue detectors at similar wavelengths. Photon-counting PMTs are therefore strongly encouraged in applications with low sample brightness. If your spectral region of interest lies in the MIR, an analogue detector is the only option. It should be paired with a powerful light source, such as a high-power laser, to offset the detector’s low yield by providing more excitation photons. 

Photons/s required to obtain a signal-to-noise ratio equal to 1 for the FLS1000 PMT detector options up to 1700 nm.

Figure 2. Photons/s required to obtain a signal-to-noise ratio equal to 1 for the FLS1000 PMT detector options up to 1700 nm.

 

Photons/s required to obtain a signal-to-noise ratio equal to 1 for the FLS1000 PMT detector options up to 1700 nm.

Figure 3. Photons/s required to obtain a signal-to-noise ratio equal to 1 for the FLS1000 analogue detector options up to 5.5 μm.

 

Are You Performing Quantum Yield Measurements?

Photoluminescence Quantum Yield (PLQY) measurements require capturing both the absorption and emission regions. In most cases, these regions are close together, allowing measurements to be made with a single detector. However, some materials have absorption and emission ranges that are far apart. In such cases, two detectors are necessary: one to measure the absorption peak and another to measure the emission peak. For example, a suitable detector combination is PMT-900/980 paired with NIR PMT-1700-LN2. This setup works well for samples that absorb in the UV or visible spectrum (~350 nm – 600 nm) and emit in the NIR (~1000 nm – 1650 nm), such as rare-earth-based materials. The 500 nm – 980 nm range is the overlapping region, as Edinburgh Instruments can supply correction files for the LN2 NIR PMT detectors starting at 500 nm. Similarly, for PLQY measurements requiring an analogue detector, such as an InGaAs, correction files are available from 600 nm onwards. Table 3 summarises all the visible and NIR detector combinations suitable for PLQY measurements. 

Table 3. Summary of the possible and non-possible combinations of visible and NIR detectors for PLQY measurements.

DetectorPMT-1400-LN2PMT-1400-TEPMT-1700-LN2PMT-1700-TEInGaAs-1650-TEInGaAs-2050-TEInGaAs-2550-TE
PMT-900PossibleNot possiblePossibleNot possiblePossiblePossiblePossible
PMT-980PossibleNot possiblePossibleNot possiblePossiblePossiblePossible
PMT-1010PossiblePossiblePossiblePossiblePossiblePossiblePossible

Note that the PMT-1010 is not recommended as the only visible PMT in a system, as its QE is lower than that of a PMT-900/980. However, it is the ideal choice for upconversion PLQY at 980 nm excitation, enabling measurement of both the sample’s excitation and emission peaks. 

 

Are You Performing Time-Resolved Measurements? 

If you require time-resolved photoluminescence (TRPL), then the detector’s time response becomes essential. The shortest measurable lifetime depends on the Instrument Response Function (IRF), which describes the instrument’s temporal response. The IRF depends on several factors, including the pulse width of the source, the detector’s temporal response, the electronics’ timing jitter, which is typically around 20 ps, and the monochromator’s dispersion, which becomes noticeable only in the picosecond time regime. These factors can be summarised in Eq. 3 as: 

FWHMdetector refers to the detector’s temporal resolution. In PMTs, this is characterised by the transit time spread (TTS). Transit time is the time it takes for photoelectrons emitted from the PMT’s photocathode to reach the anode after being multiplied by the dynodes. Each photoelectron has a unique transit time, and TTS accounts for the statistical variation in the arrival times of these electrons. A lower TTS indicates better time resolution, enabling the measurement of shorter lifetimes. For example, the standard PMT-900 has a TTS of 600 ps, whereas the high-speed hybrid photodetector (HS-HPD) has a TTS of approximately 40 ps.  

Fast detectors are often less sensitive than standard PMTs and are therefore used only for lifetime measurements. In the time-correlated single-photon counting method (TCSPC), the theoretical minimum measurable lifetime can be approximated by one-tenth of the FWHMIRF, whereas a more practical minimum estimate is given by one-third of the FWHMIRF, as: 

While Eq. 4 is the most frequently quoted minimum achievable lifetime, it is valid only under ideal experimental conditions: the sample must be a solution, and its lifetime decay curve should consist of a single exponential component. One-third of the FWHMIRF (Eq. 5) is, therefore, a more practical estimate, as it improves fit reliability when measuring not only solutions with a single exponential component but also solid-state samples with multiexponential lifetime decays. More details are available in our technical note which focuses on minimum measurable lifetimes in the FLS1000. 

Analogue detectors, as non-photon-counting devices, produce a voltage signal proportional to the incident PL intensity. To record these rapid analogue signals (transients) over time, a fast oscilloscope is required for TRPL. Edinburgh Instruments has developed TDS electronics that simplify NIR/MIR analogue detector measurements, eliminating the need for an external oscilloscope. Therefore, TDS detectors are recommended for their ease of use. However, PL lifetimes shorter than 1 μs require the external oscilloscope option, which offers higher time resolution. Oscilloscope detectors are ideally paired with high-power nanosecond lasers such as Nd:YAG lasers. 

Quick Reference Guide: PMT Detector Options 

Tables 4 and 5 summarise the spectral and temporal characteristics, as well as the typical application fields, of the available PMT detectors for the FLS1000. 

Table 4. Summary of the parameters of the available PMT detector options for the FLS1000.

DetectorSpectral Range (nm)CoolingTTSTypical Applications
PMT-900200–870Uncooled or Thermoelectric600 psGeneral PL, TRPL in the visible range
PMT-980200–950Thermoelectric600 psGeneral PL, TRPL in the visible range
PMT-1010200-1010Thermoelectric800 psPLQY for upconversion materials
PMT-1400-TE950–1400Thermoelectric400 psHighest sensitivity NIR PL, singlet oxygen measurements
PMT-1400-LN2500-1400Liquid Nitrogen800 psHighest sensitivity NIR PL
PMT-1700-TE950-1700Thermoelectric400 psHigh sensitivity deep NIR detection
PMT-1700-LN2500–1700Liquid Nitrogen800 psHigh sensitivity deep NIR detection, visible detection
High-Speed PMT230-850Thermoelectric180 psFast decays, narrow IRF
High-Speed Hybrid Photodetector220-870Thermoelectric< 40 psUltrafast dynamics, narrowest IRF

Table 5. Summary of the available analogue detector options for the FLS1000.

DetectorSpectral Range (nm)CoolingIRF
InGaAs-1650-TE870 nm – 1650 nmThermoelectric100 ns
InGaAs-2050-TE900 nm – 2050 nmThermoelectric100 ns
InGaAs-2550-TE900 nm – 2550 nmThermoelectric150 ns
InAs-3100-LN21200 nm – 3100 nmLiquid Nitrogen~150 ns
InSb-5500-LN21500 nm – 5500 nmLiquid Nitrogen500 ns

Conclusion 

Choosing the right detector for the FLS1000 involves: 

  • Matching the spectral range to your sample. 
  • Understanding brightness and selecting detectors based on sensitivity. 
  • Ensuring detector overlap for PLQY if needed. 
  • Selecting the correct time response for time-resolved measurements. 

Edinburgh Instruments provides expert guidance on selecting detectors that match gratings, sources, and measurement objectives. Contact us to finalise your system configuration and ensure optimal performance. 

RELATED PRODUCTS

FLS1000

Photoluminescence Spectrometer

VIEW

FS5

Spectrofluorometer

VIEW

Contact our expert team today to find out more about how our products can improve your research

Contact Us
PrevPrevious
Previous Automated High-throughput Upconversion Nanoparticle Spectral & Lifetime Characterisation

RESOURCES

FLS1000
Tags:
  • Technical Notes
  • FLS1000
  • FS5
  • Photoluminescence
Technical Note
Download PDF
Suggested Reading:

Measuring Fluorescence and Phosphorescence Spectra at Low Temperature Using the FLS1000 Photoluminescence Spectrometer

View more »

Benefits of Using a Gated PMT Detector in the FLS1000

View more »
Share:

Keep up to date with the latest from Edinburgh Instruments

Join our mailing list and keep up with our latest videos, app notes and more!

LOCATION:
  • Edinburgh Instruments Ltd.
    2 Bain Square, Kirkton Campus, Livingston, EH54 7DQ.
  • sales@edinst.com
  • +44 1506 425 300
ABOUT:
  • About Edinburgh Instruments
  • Techniques
  • Applications
  • Knowledgebase
  • About Edinburgh Instruments
  • Techniques
  • Applications
  • Knowledgebase
PRODUCTS:
  • Fluorescence Spectrometers
  • Raman Microscopes
  • UV-Vis Spectrophotometers
  • Transient Absorption
  • FTIR Spectrometers
  • Lasers and LEDs
  • Customisation Options
  • Software
  • Upgrades
  • All Products
  • Fluorescence Spectrometers
  • Raman Microscopes
  • UV-Vis Spectrophotometers
  • Transient Absorption
  • FTIR Spectrometers
  • Lasers and LEDs
  • Customisation Options
  • Software
  • Upgrades
  • All Products
LEGALS:
  • News
  • Events
  • Blog
  • Careers
  • Distribution & Agents
  • Contact Us
  • Terms and Conditions
  • Privacy Policy
  • News
  • Events
  • Blog
  • Careers
  • Distribution & Agents
  • Contact Us
  • Terms and Conditions
  • Privacy Policy
SOCIALS:
Youtube Linkedin X-twitter Facebook
©2024 Edinburgh Instruments. Registered in England and Wales No: 962331. VAT No: GB 271 7379 37
Manage Consent
To provide the best experiences, we use technologies like cookies to store and/or access device information. Consenting to these technologies will allow us to process data such as browsing behaviour or unique IDs on this site. Not consenting or withdrawing consent, may adversely affect certain features and functions.
Functional Always active
The technical storage or access is strictly necessary for the legitimate purpose of enabling the use of a specific service explicitly requested by the subscriber or user, or for the sole purpose of carrying out the transmission of a communication over an electronic communications network.
Preferences
The technical storage or access is necessary for the legitimate purpose of storing preferences that are not requested by the subscriber or user.
Statistics
The technical storage or access that is used exclusively for statistical purposes. The technical storage or access that is used exclusively for anonymous statistical purposes. Without a subpoena, voluntary compliance on the part of your Internet Service Provider, or additional records from a third party, information stored or retrieved for this purpose alone cannot usually be used to identify you.
Marketing
The technical storage or access is required to create user profiles to send advertising, or to track the user on a website or across several websites for similar marketing purposes.
  • Manage options
  • Manage services
  • Manage {vendor_count} vendors
  • Read more about these purposes
View preferences
  • {title}
  • {title}
  • {title}