Skip to content
Skip to content
  • News
  • Events
  • eBooks
  • Blog
  • Careers
  • Contact
  • News
  • Events
  • eBooks
  • Blog
  • Careers
  • Contact
KNOWLEDGEBASE
  • About Us
  • Products

    Fluorescence Spectrometers

    • FLS1000 Photoluminescence Spectrometer
    • FS5 Spectrofluorometer
    • LifeSpec II Lifetime Spectrometer
    • Mini-tau Lifetime Spectrometer

    Raman Microscopes

    • RM5 Raman Microscope
    • RMS1000 Multimodal Confocal Microscope

    Transient Absorption

    • LP980 Transient Absorption Spectrometer

    FTIR Spectrometers

    • IR5 FTIR Spectrometer

    Lasers and LEDs

    • Pulsed Lasers
    • Gas Lasers
    • Customisation Options
    View All Products
  • Techniques
  • Applications
KNOWLEDGEBASE
Edit Content
  • About Us
  • Products
  • Techniques
  • Applications
  • Knowledgebase
  • eBooks
  • News
  • Events
  • Blog
  • Careers
  • Contact Us

RESOURCES

TCSPC – What is Time-Correlated Single Photon Counting?

  • June 28, 2023

KEY POINTS

  • TCSPC records the time between an excitation pulse and a single emitted photon.
  • A detector converts a single photon to an electronic pulse and the TCSPC electronics accurately measure its arrival time with respect to the excitation pulse.
  • The process may be thought of like a stopwatch, each emitted sample photon has its arrival time noted.
Edit Content

Time-correlated single-photon counting (TCSPC) is the method of choice for quantifying fluorescence lifetimes. It can be used for the spectroscopic characterisation of a wide variety of solid, liquid, and gas samples, and when combined with optical microscopy it allows for fluorescence lifetime imaging (FLIM) to be performed.

  • Many thousands of arrival times are iteratively recorded and sorted by arrival time into a histogram showing their statistical intensity at a given time, as shown in Figure 1.

Simplified TCSPC Electronics Diagram

Figure 1. Simplified overview of TCSPC with two inputs (Reference START and Sample STOP) producing a histogram of timed events binned into channels split evenly across the time range.

  • In TCSPC a pulsed laser is used as the excitation source and a reference pulse from this laser is used to START the counting electronics. An emitted photon from the sample used as the STOP signal as shown in Figure 2.

Pulse Collection In TCSPC
Figure 2. Pulse collection in TCSPC where the START line is supplied from the excitation source and the STOP is connected to the sample emission. Where no STOP signal is received within the window, the electronics must reset before the next START pulse

  • A maximum of one STOP photon can be detected after each START pulse. Lasers with high repetition rates in the MHz are typically used for TCSPC in order to record millions of START-STOP times as quickly as possible.

 

TCSPC Decay Example

An example of a single exponential fluorescence decay acquired using TCSPC is shown in Figure 3. As well as the sample decay, the plot also includes the Instrument Response Function (IRF) which is characteristic of the specific measurement system, and a reconvolution fit of the decay to two components.

Fluorescence Decay Graph

Figure 3. Fluorescence decay of a sample (blue) and the IRF (red) plotted on a logarithmic scale with the fitting of the data by revolution analysis (grey) and the residuals of this fit below (black).

TCSPC can be used to measure lifetimes as low as a few picoseconds up to hundreds of nanoseconds- over 7 orders of magnitude.  At the lower limits of detection, the temporal response of the specific system, IRF, must be considered. The IRF is found by measuring the full width half maximum (FWHM) of a scattering solution at the excitation wavelength. This represents the shortest decay a system could resolve, as any fluorophore with a lifetime significantly shorter than the FWHMIRF will appear as the IRF. When recording decays on the order of magnitude of the IRF, reconvolution fitting should be used to extract the influence of the IRF on the shape of the decay function. The IRF will be affected by four components of the system: the pulse width of the excitation source used, the width response of the detector, the electronic jitter of the counting electronics, and the temporal dispersion through the emission monochromator. The response of the detector is usually the limiting factor in the sensitivity of a TCSPC system.

There must be sufficient time range to allow the fluorescence signal to decay down to the level of the background counts before the next counting cycle may begin as shown in Figure 3. Where this is not implemented, the decay will start to ‘wrap-around’ the window, distorting the peak of the next decay as it will appear to have arisen from the proceeding start event. In measurements with low sample count rates and high background and/or dark counts, the rising baseline will limit the dynamic range by masking the tail of the decay which leads to an underestimate of the lifetime.

 

Pulse Pile-up

In TCSPC a maximum of one photon can be detected after each START pulse. In Figure 4 we can see that that three photons arrive within the first START window; but after the first photon is detected, the electronics cannot see any subsequent photons. Since faster photons will arrive at the detector first, the photon histogram will be distorted towards these faster arriving photons.

Illustration of Pulse Pile-up

Figure 4. Illustration of pulse pile-up effect. Of three emitted photons in the first pulse period, only the first to arrive is seen by the electronics

This phenomenon is called pulse pile-up and would distort the measured fluorescence decay. To prevent pulse pile-up the STOP rate is limited to below 10 % of the START, to so that the statistical probability of two photons arriving in the same START window is negligibly small. This maximum STOP rate is suggested between 1 and 10 %,1–4 Edinburgh Instruments Fluoracle advises a maximum of 5 %. For instance, the repetition rate of a 100 kHz source would have recommended maximum count rate of 5,000 Hz.

Operating Modes in TCSPC

The examples described here all show a system operating in the conventional Forward Mode. The drawback of Forward Mode is observed when using excitation sources with high repetition rates, like Ti:Sapphire lasers which can reach well over 50 MHz. This is because the time between START pulses is less than the time required to reset the electronics, known as the dead time. If START pulses arrive during the dead time the electronics will not start so higher START rates will not decrease the decay acquisition time and photosensitive samples will be more prone to bleaching due to increased exposure. Instead, the system may be operated at the rate of the emission pulses by using Reverse Mode. In Reverse Mode, an emitted photon from the sample is used to START counting, and the excitation laser pulse STOPS the counting. Starting the counting electronics from the receipt of an emitted photon means that the counting electronics will only be activated when there is a count to be recorded. Reverse Mode has great utility in FLIM applications, where a large area must be scanned, and the decays can be acquired as quickly as possible.

 

TCSPC Electronics

The main components of the TCSPC electronics are the Constant Fraction Discriminator (CFD), an electronic delay (Delay), the Time-to-Amplitude Converter (TAC), an amplifier (not shown), an Analogue-to-Digital Converter (ADC), and the digital memory (MEM) shown in Figure 5.

Main Components of electronics for Time Correlated Single Photon Counting

Figure 5. Schematic showing the main components of signal processing in TCSPC electronics.

START and STOP signals are received at the constant fraction discriminator (CFD) where the pulse must surpass the Threshold voltage. This eliminates low amplitude noise from being counted as a real signal. The zero crossing may be adjusted to tens of mV above or below zero to reduce the timing jitter due to irregular pulse height.3 This produces an electronic pulse which is independent of amplitude to be passed to the TAC. The STOP signal may be delayed to position it optimally in the recorded window so that the background counts can be viewed before the decay, and to confirm the emission drops back to the background level by the end of the window as shown in Figure 3.

The TAC starts ‘counting’ (gain) until a STOP signal (emitted photon) is received. If no signal is received in this time, the charge is dispersed to reset the TAC for the next cycle. When a signal is received, the amplitude which has accumulated in this time is passed to the amplitude to digital converter (ADC) and is binned to the appropriate channel, building up a histogram of time against counts. The resolution of the channels may be altered by selecting a shorter time range or increasing the number of channels.

The time window can be tuned to detect fluorophores on a time scale from 10 picoseconds to 8 microseconds depending on the system configuration.2 Longer photoluminescence decay times can be measured using Multi-Channel Scaling (MCS).

 

References

1            W. (Wolfgang) Becker, Advanced time-correlated single photon counting techniques, Springer, 2005.

2            J. R. Lakowicz, Principles of Fluorescence Spectroscopy Third Edition, 2010.

3            W. Becker, The bh TCSPC Handbook, 8th edn., 2019.

4            B. Valeur and M. N. Berberan-Santos, Molecular Fluorescence, Wiley-VCH, 2nd edn., 2012.

 

If you have enjoyed this blog, be sure to subscribe to our enewsletter and follow us on social media to keep up to date with the latest products news and updates.

RELATED PRODUCTS

RMS1000

Multimodal Confocal Microscope

VIEW

FLS1000

Photoluminescence Spectrometer

VIEW

EPL Series

Pulsed Lasers

VIEW

Contact our expert team today to find out more about how our products can improve your research

Contact Us
PrevPrevious
NextNext
Previous Instrument Response Function Widths & Minimum Measurable Lifetimes in the FLS1000 Next Near Infrared PMT Options for the FLS1000 Photoluminescence Spectrometer

RESOURCES

Tags:
  • Spectral School
  • FLS1000
  • FS5
  • FLIM
  • Photoluminescence
Suggested Reading:

Instrument Response Function Widths & Minimum Measurable Lifetimes in the FLS1000

View more »

What is Fluorescence Lifetime?

View more »
Share:

Keep up to date with the latest from Edinburgh Instruments

Join our mailing list and keep up with our latest videos, app notes and more!

LOCATION:
  • Edinburgh Instruments Ltd.
    2 Bain Square, Kirkton Campus, Livingston, EH54 7DQ.
  • sales@edinst.com
  • +44 1506 425 300
ABOUT:
  • About Us
  • Techniques
  • Applications
  • Knowledgebase
  • About Us
  • Techniques
  • Applications
  • Knowledgebase
PRODUCTS:
  • Fluorescence Spectrometers
  • Raman Microscopes
  • UV-Vis Spectrophotometers
  • Transient Absorption
  • FTIR Spectrometers
  • Lasers and LEDs
  • Customisation Options
  • Software
  • Upgrades
  • All Products
  • Fluorescence Spectrometers
  • Raman Microscopes
  • UV-Vis Spectrophotometers
  • Transient Absorption
  • FTIR Spectrometers
  • Lasers and LEDs
  • Customisation Options
  • Software
  • Upgrades
  • All Products
LEGALS:
  • News
  • Events
  • Blog
  • Careers
  • Contact Us
  • Terms and Conditions
  • Privacy Policy
  • News
  • Events
  • Blog
  • Careers
  • Contact Us
  • Terms and Conditions
  • Privacy Policy
SOCIALS:
Youtube Linkedin X-twitter Facebook
©2024 Edinburgh Instruments. Registered in England and Wales No: 962331. VAT No: GB 271 7379 37
Manage Consent
To provide the best experiences, we use technologies like cookies to store and/or access device information. Consenting to these technologies will allow us to process data such as browsing behaviour or unique IDs on this site. Not consenting or withdrawing consent, may adversely affect certain features and functions.
Functional Always active
The technical storage or access is strictly necessary for the legitimate purpose of enabling the use of a specific service explicitly requested by the subscriber or user, or for the sole purpose of carrying out the transmission of a communication over an electronic communications network.
Preferences
The technical storage or access is necessary for the legitimate purpose of storing preferences that are not requested by the subscriber or user.
Statistics
The technical storage or access that is used exclusively for statistical purposes. The technical storage or access that is used exclusively for anonymous statistical purposes. Without a subpoena, voluntary compliance on the part of your Internet Service Provider, or additional records from a third party, information stored or retrieved for this purpose alone cannot usually be used to identify you.
Marketing
The technical storage or access is required to create user profiles to send advertising, or to track the user on a website or across several websites for similar marketing purposes.
Manage options Manage services Manage {vendor_count} vendors Read more about these purposes
View preferences
{title} {title} {title}