Skip to content
Skip to content
  • News
  • Events
  • eBooks
  • Blog
  • Careers
  • Contact
  • News
  • Events
  • eBooks
  • Blog
  • Careers
  • Contact
KNOWLEDGEBASE
  • About Us
  • Products

    Fluorescence Spectrometers

    • FLS1000 Photoluminescence Spectrometer
    • FS5 Spectrofluorometer
    • LifeSpec II Lifetime Spectrometer
    • Mini-tau Lifetime Spectrometer

    Raman Microscopes

    • RM5 Raman Microscope
    • RMS1000 Multimodal Confocal Microscope

    Transient Absorption

    • LP980 Transient Absorption Spectrometer

    FTIR Spectrometers

    • IR5 FTIR Spectrometer

    Lasers and LEDs

    • Pulsed Lasers
    • Gas Lasers
    • Customisation Options
    View All Products
  • Techniques
  • Applications
KNOWLEDGEBASE
Edit Content
  • About Us
  • Products
  • Techniques
  • Applications
  • Knowledgebase
  • eBooks
  • News
  • Events
  • Blog
  • Careers
  • Contact Us

RESOURCES

Understanding Bandpass and Resolution in Photoluminescence Spectroscopy

  • August 21, 2025

KEY POINTS

  • Bandpass defines the range of the wavelength distribution that a spectrometer allows to pass. Related to this bandwidth quantifies the width of this pass window.
  • Spectral resolution measures its ability to distinguish between very close wavelengths, revealing fine spectral details.
  • Components like slits, gratings, monochromators, and detectors all play a vital role in determining the achievable spectral resolution.
Edit Content

What are Bandpass and Spectral Resolution? 

In photoluminescence spectroscopy, two key concepts frequently arise when discussing how a spectrometer detects light: bandpass and spectral resolution. While related, they describe distinct capabilities of your instrument. Together, they define the ability of an instrument to distinguish between spectral features. 

The bandpass refers to the specific range of wavelengths that an instrument is designed to isolate. Think of it as the “window” of the electromagnetic spectrum that’s allowed to pass through to the detector. Related to bandpass, bandwidth quantifies the width of this window. 

On the other hand, spectral resolution describes the wavelength difference between two lines in a spectrum that can still be clearly distinguished. Essentially, it tells you how finely a spectrometer can “see” detail in a spectrum. Higher spectral resolution means the instrument can pick out narrower spectral features. These terms are summarised in Table 1. 

 

Table 1. Relationship between bandpass, bandwidth and spectral resolution. 

Term  Definition  Relation to Others 
Bandpass  The specific range of wavelengths a spectrometer is designed to transmit, defining the “window” of light reaching the detector (e.g. bandpass of 400 – 410 nm).  Bandpass is the wavelength interval an instrument isolates, characterised by its centre wavelength (CWL) and bandwidth. A narrower bandpass implies higher spectral resolution. 
Bandwidth  The width of the bandpass, quantifying the specific range of wavelengths the spectrometer collects (e.g. bandwidth of 10 nm).  A quantitative measure of a spectrometer’s bandpass. A smaller bandwidth indicates finer spectral resolution. 
Spectral Resolution  The ability of a spectrometer to distinguish between different wavelengths, referring to how narrow and numerous the detectable spectral bands are.  Directly related to bandwidth. Higher spectral resolution means the spectrometer detects more and narrower bands, which means smaller bandwidths for each band, providing more detailed spectral information. 

 

Why is Resolution Important in Photoluminescence?

Although spectral resolution is crucial in Raman spectroscopy, its importance in fluorescence varies depending on the fluorophore’s spectral profile. This is because many fluorophores exhibit broad, relatively featureless emission spectra. So, in many cases, changing spectral resolution won’t noticeably affect the observed fluorescence spectrum. 

However, adequate (or even high) wavelength resolution becomes critically important in fluorescence in specific scenarios: 

  • When you are analysing fluorophores with structured emission – those that naturally exhibit distinct, narrow peaks or vibrational fine structure. 

 

  • When you are trying to identify subtle changes in a molecule’s environment, which might cause a slight shift in its fluorescence emission wavelength.

 

  • When you need to resolve multiple, overlapping emission spectra from different fluorophores within the same sample.  

 

Key Factors Influencing Spectral Resolution

Several key spectrometer components directly impact its bandpass and spectral resolution. These include the slit width, grating dispersion, monochromator design and detector configuration (Figure 1). 

Figure 1. Key Factors Influencing Spectral Resolution. 

 

  1. Slit Width

In a spectrometer, slits are crucial. These adjustable apertures precisely control the amount of light entering and exiting the monochromator, directly impacting both the signal you measure and the detail you can see. 

The slit width is what controls the spectrometer’s bandwidth (the width of the bandpass). For example, setting an emission wavelength of 500 nm with a 2 nm bandwidth means the slits and monochromator are positioned to collect light spanning from 499 nm to 501 nm (500 nm ± 1 nm). A narrower bandwidth signifies more precise wavelength selection and thus better spectral resolution. 

It’s important to distinguish between the physical slit width and the spectral bandwidth. The spectral bandwidth, often defined as the peak width at half height of the light exiting the monochromator, will always be less than the physical slit width.  

The intensity of light passing through the monochromator is roughly proportional to the square of the slit width. Wider slits allow significantly more light through, boosting the signal-to-noise ratio. Conversely, narrower slits improve resolution but drastically reduce light intensity, potentially leading to noisier data. 

 

Calculating Bandpass from Slit Width: 

The spectrometer’s bandpass can be calculated using Equation 1. 

Here, slit width usually refers to the larger of the entrance or exit slit width. The Reciprocal Linear Dispersion (RDL) is a property of your spectrometer’s grating and optics that describes how well a dispersive element, separates different wavelengths of light spatially. It is defined as the change in wavelength per unit distance along the spectrum at the monochromators exit focal plane.  

 

  1. Diffraction Gratings Dispersion

Diffraction gratings, found in both excitation and emission monochromators, are fundamental to separating light into its constituent wavelengths. 

Gratings are available with different groove densities (grooves/mm). A grating with higher groove density gives higher reciprocal dispersion, meaning it spreads the light more effectively (Figure 2). Diffracting the light wider provides improved spectral resolution. For instance, a monochromator’s bandpass with a 1200 grooves/mm grating can be half that of a 600 grooves/mm grating (assuming the same slit width). However, a larger dispersion of light comes at the cost of spectral intensity. 

Figure 2. Effect of Groove Density: A higher groove density spreads light more effectively, improving resolution. A lower groove density spreads light less, reducing resolution. 

 

A grating’s resolving power is fundamentally related to its ability to separate wavelengths and is dependent on the spectral order and the number of grooves under illumination (Equation 2).  

Where: 

  • R = resolving power, indicating how well the grating can distinguish between closely spaced wavelengths 

 

  • λ = average wavelength of light being diffracted. 

 

  • Δλ = the smallest difference in wavelength that the grating can resolve. 

 

  • m = the order of the diffraction, where diffracted angles m = 1 are considered to be 1st order diffraction, m = 2 are considered to be 2nd order diffraction, and so on.  

 

  • N = the total number of illuminated grooves on the grating. 

 

Often, there are so many grooves under illumination that the entrance and exit slits become the limiting factor for the overall system resolution, rather than the grating’s theoretical resolving power. 

 

  1. Monochromator Design

The monochromator is the entire optical system comprising the slits, mirrors and the grating, responsible for selecting a specific wavelength of light. Its design significantly impacts spectral resolution. 

The focal length of the monochromator affects the dispersion of light. A longer focal length leads to increased dispersion of light, thereby enhancing resolution. In addition, the specific arrangement of mirrors and gratings within the monochromator (e.g. Czerny-Turner) influences optical aberrations and overall resolution. More complex designs (like double monochromators) can offer significantly improved resolution and reduced stray light, which is unwanted light that reaches the detector without passing through the intended optical path. Stray light can obscure weak fluorescence signals, effectively reducing overall resolution. 

 

  1. Detector Configuration

The detector converts the light signal into an electrical signal, and its characteristics contribute to the overall resolution. The spectrometer design and detector type can determine the optimal slit setup, for example: 

  • Scanning Monochromators with PMT Detector: In these systems, where a single detector is used and the grating rotates to scan wavelengths, the entrance slit of the emission monochromator primarily dictates the resolution. If the exit slit is the same size or smaller than the image of the entrance slit, it has little further impact on resolution. Therefore, matching the exit slit size to the entrance slit is common to maximize light throughput without compromising resolution. 

 

  • Diode Array Spectrometers: These instruments use a fixed grating and a linear array of detectors (like a CCD camera). Here, the individual detector elements (pixels) act as the “exit slits.” These pixels are typically much smaller than the entrance slit, and their size plays a significant role in defining the effective resolution of the system. Smaller, denser pixels allow for finer sampling of the dispersed light, leading to better effective resolution. 

While not directly affecting resolution, a detector’s quantum efficiency (QE -its efficiency in converting photons to electrons) impacts the signal-to-noise ratio. A higher QE allows for better detection of weak signals, which is crucial when operating at high resolution where signal intensity is often reduced. 

Furthermore, inherent detector noise can limit the ability to distinguish between closely spaced, low-intensity peaks, thereby reducing practical resolution even if the optical components are theoretically capable of higher performance.  

 

How is Spectral Resolution Measured?  

A function that defines the spectral resolution of your system and determines how accurately you can resolve narrow spectral features, is the Instrument Line Profile (ILP). The ILP reflects the shape of the spectrum produced if it was illuminated with perfectly monochromatic light (Figure 3). For example, if you were to analyse a light source with a single monochromatic wavelength (λx) on a ‘perfect’ spectrometer the resulting spectrum should be a perfect line at precisely λx. In reality, you would observe spectral broadening resulting in a line profile with finite width. The finite line width of the instrumental response limits the ability to distinguish lines that are close to each other. 

Figure 3. The ILP determines how accurately you can resolve narrow spectral features. A). Theoretical emission spectrum of single monochromatic wavelength source. B). ILP of monochromatic source. 

The recorded spectrum is mathematically described as a convolution of the real spectrum of your sample and the ILP (Equation 3). 

 

The ILP is a function of various parameters including: 

  • Monochromator entrance slit width  

 

  • Monochromator exit slit width  

 

  • Diffraction phenomena 

 

  • Optical aberrations 

 

  • Quality of system components and alignment. 

 

What is the Full Width Half Maximum (FWHM)? 

Spectral resolution is measured by using the Full Width Half Maximum (FWHM) of narrow monochromatic light through the spectrometer (Figure 4). The FWHM determines the sharpness or broadening of a peak; a lower FWHM indicates a narrower peak and, consequently, higher spectral resolution.  

The spectral bandpass of an instrument is also defined by the FWHM of the wavelength distribution passed by the exit slit. 

Figure 4. Full Width Half Maximum (FWHM). 

 

The steps for calculating the FWHM are: 

  1. Identify the maximum value of the emission (fmax). This is the highest point of the peak. 
  2. Calculate half of this maximum value
  3. Find the two curve points that correspond to this half maximum value (at x1 and x2)
  4. Measure the width between these points (x2 – x1). This is the FWHM. 

 

Calculating the FWHM of the Instrumental Profile 

The overall resolution of an instrument is limited by the FWHM of its instrumental profile, which is dependent on various causes of line broadening. The total FWHM is calculated using the Equation 4. 

 

Where: 

  • dλ2(slits) represents the bandpass determined by finite spectrometer slit widths and linear dispersion of the grating. 

 

  • dλ2(resolution) encompasses the limiting resolution of the spectrometer which incorporates system aberrations, diffraction effects and the laser line width of the system. 

 

  • dλ2(line) represents the natural line width of the spectral line being probed. 

 

Example: Bandwidth’s Effect on Spectral Resolution 

As discussed, bandwidth significantly impacts spectral resolution. To illustrate this, let’s consider an example using a sample of Nd:YAG (Neodymium-doped Yttrium Aluminium Garnet). 

We chose Nd:YAG because of its inherently narrow spectral features, typically around 2 nm wide, making it an ideal candidate for demonstrating the crucial role of good spectral resolution. The collected emission spectra, obtained using varying emission bandwidths, are shown in Figure 5. 

 

Figure 5. Emission spectra of Nd:YAG collected on FLS1000 PL Spectrometer. A). Bandwidths 0.5, 1, 2, 3, 4 and 5 nm. B). Bandwidths 0.5 nm versus 5 nm. Experimental parameters: Detector = PMT-1400-TE, Excitation = Xenon Lamp, Emission grating = 600 grooves/mm, λex = 355 nm, Step size = 0.2 nm, Dwell Time = 1s. 

 

The spectra clearly demonstrate that as the bandwidth is increased, spectral resolution is progressively lost. The sharp peaks observed with a 0.5 nm bandwidth, for instance, become significantly broadened when the bandwidth is increased to 5 nm. This occurs because a larger bandwidth allows a wider spectral range to pass through the monochromator simultaneously. With a broader spectral range, the instrument’s ability to distinguish individual features decreases, leading to a loss of fine spectral detail. 

 

The Crucial Trade-Off: Signal versus Resolution 

Ultimately, when measuring fluorescence, there is a choice: do you prioritise high resolution or strong signal intensity? 

  • To achieve higher resolution, you typically need to use narrower slits or gratings with higher groove densities, which means less light reaches the detector. This inevitably results in a weaker signal. 

 

  • Conversely, to achieve a stronger signal, you might widen the slits, which lets in more light but inherently lowers your spectral resolution. 

Finding the right balance between these two factors is paramount for obtaining meaningful and accurate fluorescence data for your specific experimental goals.  

 

Conclusion 

Bandpass and spectral resolution are key parameters in fluorescence spectroscopy, directly affecting the ability to resolve fine spectral features. While many fluorophores emit broad spectra, high resolution becomes essential when distinguishing overlapping signals, detecting subtle shifts, or analysing structured emissions. 

Spectral resolution is shaped by slit width, grating dispersion, monochromator design, and detector configuration—each involving a trade-off between signal strength and resolution. Using FWHM and understanding the instrument line profile helps optimise this balance. By carefully adjusting these variables, users can achieve accurate, high-resolution fluorescence data tailored to their experimental needs. 

RELATED PRODUCTS

FLS1000

Photoluminescence Spectrometer

VIEW

FS5

Spectrofluorometer

VIEW

Contact our expert team today to find out more about how our products can improve your research

Contact Us
PrevPrevious
Previous Photo- and Radioluminescence of Molybdenum Nanoclusters for Photodynamic Therapy

RESOURCES

Tags:
  • Spectral School
  • Photoluminescence
Suggested Reading:

Second Order Diffraction through a Monochromator – Common Errors in Fluorescence Spectroscopy

View more »

Spectral Resolution in Raman Spectroscopy

View more »

A Short Introduction to Fluorescence Measurements & Instrumentation

View more »
Share:

Keep up to date with the latest from Edinburgh Instruments

Join our mailing list and keep up with our latest videos, app notes and more!

LOCATION:
  • Edinburgh Instruments Ltd.
    2 Bain Square, Kirkton Campus, Livingston, EH54 7DQ.
  • sales@edinst.com
  • +44 1506 425 300
ABOUT:
  • About Us
  • Techniques
  • Applications
  • Knowledgebase
  • About Us
  • Techniques
  • Applications
  • Knowledgebase
PRODUCTS:
  • Fluorescence Spectrometers
  • Raman Microscopes
  • UV-Vis Spectrophotometers
  • Transient Absorption
  • FTIR Spectrometers
  • Lasers and LEDs
  • Customisation Options
  • Software
  • Upgrades
  • All Products
  • Fluorescence Spectrometers
  • Raman Microscopes
  • UV-Vis Spectrophotometers
  • Transient Absorption
  • FTIR Spectrometers
  • Lasers and LEDs
  • Customisation Options
  • Software
  • Upgrades
  • All Products
LEGALS:
  • News
  • Events
  • Blog
  • Careers
  • Contact Us
  • Terms and Conditions
  • Privacy Policy
  • News
  • Events
  • Blog
  • Careers
  • Contact Us
  • Terms and Conditions
  • Privacy Policy
SOCIALS:
Youtube Linkedin X-twitter Facebook
©2024 Edinburgh Instruments. Registered in England and Wales No: 962331. VAT No: GB 271 7379 37
Manage Consent
To provide the best experiences, we use technologies like cookies to store and/or access device information. Consenting to these technologies will allow us to process data such as browsing behaviour or unique IDs on this site. Not consenting or withdrawing consent, may adversely affect certain features and functions.
Functional Always active
The technical storage or access is strictly necessary for the legitimate purpose of enabling the use of a specific service explicitly requested by the subscriber or user, or for the sole purpose of carrying out the transmission of a communication over an electronic communications network.
Preferences
The technical storage or access is necessary for the legitimate purpose of storing preferences that are not requested by the subscriber or user.
Statistics
The technical storage or access that is used exclusively for statistical purposes. The technical storage or access that is used exclusively for anonymous statistical purposes. Without a subpoena, voluntary compliance on the part of your Internet Service Provider, or additional records from a third party, information stored or retrieved for this purpose alone cannot usually be used to identify you.
Marketing
The technical storage or access is required to create user profiles to send advertising, or to track the user on a website or across several websites for similar marketing purposes.
Manage options Manage services Manage {vendor_count} vendors Read more about these purposes
View preferences
{title} {title} {title}