Ratiometric Fluorescence Sensor for Hypochlorite & Ascorbic Acid | MOF

Metal Organic Framework as a Ratiometric Fluorescence Sensor for Hypochlorite and Ascorbic Acid

Hypochlorite (ClO) is commonly used for disinfecting tap water, but its concentration needs to be monitored to make sure it is effective but not toxic. Researchers in Fujian, China, led by Rong Cao and Zu-Jin Lin, have developed a new method for sensing ClOemploying fluorescence spectroscopy and novel metal organic frameworks. The authors used an Edinburgh Instruments FS5 Spectrofluorometer to characterise and optimise the sensor’s response towards ClO. They then employed this fluorescence sensor to detect ascorbic acid, an essential nutrient for the human body.

Metal organic frameworks (MOFs) are compounds containing metal ions and organic ligands arranged in a regular coordination network. They are useful in sensing applications since their porous structures enable the absorption of analytes of a particular size. Moreover, it is possible to employ a fluorescent MOF as a sensor if its photoluminescence properties change in the presence of the analyte. This idea has been previously exploited for ClOdetection, but always based on a single emission intensity. In this study, Zeng et al. propose a ratiometric fluorescence sensor to improve sensitivity and robustness. In a ratiometric fluorescence sensor, the intensities of two fluorescence signals with different responses to an analyte are measured. Using the ratio of two signals reduces the error in the measurement compared to a single emission sensor. 

Results

The authors synthesised a metal organic framework containing bipyridine-based BPyDC2− and H2N- BPyDC ligands as well as Eu(III) ions. The fluorescence emission spectrum from this MOF shows two components: BPyDC2− ligands contribute with a broad emission band in the blue whilst Eu(III) presents characteristic narrow bands in the red. The authors found that Eu(III) emission was practically independent of ClO concentration, while the blue band was strongly dependent on it. Fast and accurate measurements performed in the FS5 Spectrofluorometer allowed the authors to establish a calibration curve based on the ratio between the two signals (Figure 1).

Fluorescence Sensor. Ratiometric Fluorescence: Figure 1. (a) Fluorescence emission spectra from the Eu/BPyDC MOF studied under different concentrations of hypochlorite, acquired in an FS5 Spectrofluorometer. (b) Ratiometric fluorescence calibration curve obtained from the data in (a).

Figure 1. (a) Fluorescence emission spectra from the Eu/BPyDC MOF studied under different concentrations of hypochlorite, acquired in an FS5 Spectrofluorometer. (b) Ratiometric fluorescence calibration curve obtained from the data in (a).

Zeng, et al. went a step further in their study and utilised their ClO sensor for the detection of ascorbic acid. Ascorbic acid reacts with ClOand therefore can return the sensor to its initial state. The authors found that upon addition of ascorbic acid, 471 nm fluorescence emission is recovered and it is possible to establish a calibration curve dependent on ascorbic acid concentration. Therefore, the system acts as an “on-off-on” sensor for ascorbic acid.

Further experiments helped establish the sensing mechanism, excluding energy transfer, inner filter effect, and oxidation of -NH2 groups as possible processes. Fluorescence lifetimes recorded in an Edinburgh Instruments FLS980 confirmed that the reason for the reduction in fluorescence is dynamic quenching (Figure 2). The authors hypothesise that quenching occurs due to hydrogen bonding between ClOand the -NH2 groups in the metal organic framework. Ascorbic acid reacts with ClOand gradually returns the fluorescence sensor to its initial “on” state. This is sketched in Figure 3.

Fluorescence Sensor. Ratiometric Fluorescence: Time-resolved photoluminescence decays from Eu/BPyDC MOF under different concentrations of hypochlorite, acquired in an FLS980 Photoluminescence Spectrometer.

Figure 2. Time-resolved photoluminescence decays from Eu/BPyDC MOF under different concentrations of hypochlorite, acquired in an FLS980 Photoluminescence Spectrometer.

 

Proposed mechanism for the “on-off-on” fluorescence sensor.

Figure 3. Proposed mechanism for the “on-off-on” fluorescence sensor. In the “on” state, both the pyridine-containing units and the Eu(III) ions emit fluorescence. Hypochlorite enters the MOF cavities and forms hydrogen bonds with the BPyDC units, quenching their fluorescence. Ascorbic acid removes ClO returning the sensor to its “on” state.

Conclusions

Research-grade fluorescence spectrometers such as the Edinburgh Instruments FS5 play a key part in the design and optimisation of fluorescence sensors. The FS5 offers both spectral and time-resolved measurements, which are essential for understanding the sensing mechanism. In their publication, Zeng et al. make excellent use of Edinburgh Instruments equipment for the development of a new ratiometric fluorescence sensor using metal organic frameworks.

Link to Article: https://pubs.acs.org/doi/10.1021/acs.inorgchem.9b02251

Figures reprinted with permission from Inorg. Chem. 2019, 58, 13360. Copyright © 2019, American Chemical Society.

Find out More

If you have enjoyed this application note from Edinburgh Instruments about Metal Organic Frameworks as a Ratiometric Fluorescence Sensor for Hypochlorite and Ascorbic Acid and would like to be the first to know about our future research articles, please take a moment to join us on social media and sign up to our infrequent eNewsletter via the link below.

Alternatively, if you would like to find out more information about our FS5 Spectrofluorometer, which has been used in this research, please don’t hesitate to get in touch with a member of our team today.