FL920P Spectrometer - Data Acquisition

The FL920P employs the measurement principle of Time Correlated Single Photon Counting (TCSPC) for fluorescence measurements from 100picoseconds to 10microseconds and then employs the principle of Multi-Channel scaling (MCS) for measurements from 1microseocnd to 10seconds. For TCSPC measurements we employ our TCC900 card, for MCS measurements our PCS900 fast photon counter scalar card. Both these PC plug-in cards have PCI interface. The PCS900 card can operate in two modes for both steady state and multi-channel scaling measurements. For details of MCS operation please go to the dedicated section in the web site by following this link MCS.
All important operating parameters of the PCS900 are fully computer controlled and include input threshold, number of sweeps and gate width
 

The PCS900 card features :
PCI interface
3 channel counter
Maximum count frequency of 100MHz, minimum pulse width 10ns
Up to 4000 gates of width 300ns to 0.4s
Up to 232 counts per gate
Accepts +/- 10mV to +/-5V input signal levels
Multi channel scaling capability
Accepts +1 to +5V external trigger
Plug ‘n’ Play PCI bus interface.
Standard half size PCI board dimensions

The TCC card features:
• PCI interface
• Time ranges 2.5ns to 50microseconds full scale
• Time resolution from 610fs/ch
• Ultra-low jitter <25ps
• Forward and Reverse modes
• Exceptional differential non-linearity
• Integrated shifting delays

 

 
 

 

 

 